NANOpure Diamond Analytical Water System

Highlights
  • Pure and Simple: NANOpure® DIamond™ Reagent Grade (Type 1) Water System
  • Convenient one-touch dispensing

DESCRIPTION $ QTY
ML3599 NANOpure® Diamond™ Analytical Water System - 13.5"W x 19.5"D x 17"H $4,160.00/ea
ML3600 Diamond Kit Type I R/O & Distilled Feed Cartridge - $392.00/ea
ML3601 Diamond Kit Type I Deionized Feed Cartridge - $588.00/ea
ML3602 Remote Dispenser - $620.00/ea
ML1742 Final Filter 0.2 micron - $99.00/ea
ML1743 NIST Calibrations Module - $129.00/ea
  • Eliminates carboy overflow with volumetric and timed dispense capabilities
  • Provides complete quality monitoring through measurements of resistivity and TOC
  • Simple maintenance with one-connection cartridge pack and automatic sanitization
  • Space-saving compact design
  • Tubing connections included
  • Wall- and bench-mount options (wall mount hardware included)
  • Easy-to-read display provides up to two lines of 24 characters each for accurate reading of resistivity, conductivity, temperature, and Total Organic Carbon (TOC)
  • Bright LED indicator provides at-a-glance status of system (idle, recirculation, standby, or alarm)
  • UV oxidation chamber provides organic removal and controls bacterial growth within system
  • New pump technology provides optimum flow and quiet operation
  • ML1742 0.2 micron Final Filter available
  • One-connection cartridge installs easily
  • Optional remote dispenser delivers water up to eight feet away
  • Integral pressure regulator allows feed-water pressures up to 100psig without additional equipment

This easy-to-install, easy-to-use water system is suitable for most laboratory needs including, atomic absorption (AA), graphic furnace (AA), standard preparation, and high purity rinsing.

General System Specifications
Flow Rate1.5L/min @ 10psig with a new final filter
Automatic Volumetric Dispense:
• Range0.25-60 L
• Accuracy±5% (1 L or greater)
• Repeatability± 3%
Automatic Timed Dispense1-40 minutes
Clearances:
• Sides9” for servicing
• Above3” for removal of top cover
• Front4.75” for opening front door
Electrical90-240 VAC, 100 VA, 47-63 Hz, 1 phase
 
Feed-Water Requirements
Required PretreatmentReverse osmosis, deionization, distillation
Total Organic Carbon< 1 ppm
Turbidity<1.0 NTU
PressureGravity feed to 7kj/cm (100psig)
Temperature4-40°C (40-104°F)
Total Dissolved Solids< 70 ppm as CaCO
Silica< 1 ppm
Silt Density Index< 5%
 
Product-Water Quality (Typical Values)
Resistivity 18.2 MΩ-cm (< 0.055 µS/cm) @ 25°C
TOC:
• UV and UV/UF1-5 ppb
• Analytical and UF5-10 ppb, dependent on feed-water TOC
Pyrogen:
• UF and UV/UF< 0.001 EU/mL dependent on frequency of sanitization
• Bacteria< 1 CFU/mL
• Particles (0.2 µm)<1/mL
• Silica< 3 ppb
UV Absorbence @ 254 nm< 0.0001 AU