NANOpure Diamond Analytical Water System
Highlights
- Pure and Simple: NANOpure® DIamond™ Reagent Grade (Type 1) Water System
- Convenient one-touch dispensing
Description
- Eliminates carboy overflow with volumetric and timed dispense capabilities
- Provides complete quality monitoring through measurements of resistivity and TOC
- Simple maintenance with one-connection cartridge pack and automatic sanitization
- Space-saving compact design
- Tubing connections included
- Wall- and bench-mount options (wall mount hardware included)
- Easy-to-read display provides up to two lines of 24 characters each for accurate reading of resistivity, conductivity, temperature, and Total Organic Carbon (TOC)
- Bright LED indicator provides at-a-glance status of system (idle, recirculation, standby, or alarm)
- UV oxidation chamber provides organic removal and controls bacterial growth within system
- New pump technology provides optimum flow and quiet operation
- ML1742 0.2 micron Final Filter available
- One-connection cartridge installs easily
- Optional remote dispenser delivers water up to eight feet away
- Integral pressure regulator allows feed-water pressures up to 100psig without additional equipment
This easy-to-install, easy-to-use water system is suitable for most laboratory needs including, atomic absorption (AA), graphic furnace (AA), standard preparation, and high purity rinsing.
| General System Specifications | |
| Flow Rate | 1.5L/min @ 10psig with a new final filter |
| Automatic Volumetric Dispense: | |
| • Range | 0.25-60 L |
| • Accuracy | ±5% (1 L or greater) |
| • Repeatability | ± 3% |
| Automatic Timed Dispense | 1-40 minutes |
| Clearances: | |
| • Sides | 9” for servicing |
| • Above | 3” for removal of top cover |
| • Front | 4.75” for opening front door |
| Electrical | 90-240 VAC, 100 VA, 47-63 Hz, 1 phase |
| Feed-Water Requirements | |
| Required Pretreatment | Reverse osmosis, deionization, distillation |
| Total Organic Carbon | < 1 ppm |
| Turbidity | <1.0 NTU |
| Pressure | Gravity feed to 7kj/cm (100psig) |
| Temperature | 4-40°C (40-104°F) |
| Total Dissolved Solids | < 70 ppm as CaCO |
| Silica | < 1 ppm |
| Silt Density Index | < 5% |
| Product-Water Quality (Typical Values) | |
| Resistivity | 18.2 MΩ-cm (< 0.055 µS/cm) @ 25°C |
| TOC: | |
| • UV and UV/UF | 1-5 ppb |
| • Analytical and UF | 5-10 ppb, dependent on feed-water TOC |
| Pyrogen: | |
| • UF and UV/UF | < 0.001 EU/mL dependent on frequency of sanitization |
| • Bacteria | < 1 CFU/mL |
| • Particles (0.2 µm) | <1/mL |
| • Silica | < 3 ppb |
| • UV Absorbence @ 254 nm | < 0.0001 AU |